HEPHY Vienna
Longterm multi It Setup

This measurement setup allows a simultaneous longtime dark current measurement of 10 CMS silicon strip sensors at room temperature. The measurement box is filled with nitrogen to reduce the humidity. During the measurment the temperature and the humidity are permanently monitored.

 

Updates:

Setup Schematics

This setup is used 10 times in parallel with 1 SMU and a scanning Voltmeter, which allows a parallel measurement of 40 channels.

Setup Components

Description of the Instruments:

 

Pictures of the current setup

Measurement data

This plot shows a first test measurement of the setup. Four sensors have been measured. All these sensors have a very low and constant dark current over the whole measurement time of 112 h (4.5 days).
This can only be achieved by a uniform temperature distribution (+/- 1°C) during the measurment and a decrease of the humidity in the test box.
This has been done by the help of nitrogen gas, which was filled in the box at the measurement start to lower it to 10%. Because of leaky components the humidity increases exponentially after the gas supply has stopped. It reaches 30% (which is the limit) after approx. 24 h and has to be refilled. This can be seen in the plot below.

 

 

Labview Software


last modified:
Thomas Bergauer